E.3.3 Known configuration
E.3.3.1 General
This exchange takes place when the manager returns the AssociateResult code of accepted because the manager had preiously receied and processed the configuration corresponding to the dev.config.id sent by the agent. In this ease, there is no exchange of configuration information, and the manager and agent havc moved into thc Operating slate.
E.3.3.2 Remote operation invoke event report configuration
Since the manager was already aware of the agent’s configuration, the Configuring state is skipped, and no event report invocation is generated by the agent.
E.3.3.3 Remote operation response event report configuration
The Configuring state has been skipped. No een1 report invocation is generated by the agent, so the manager does not generate any response.
E.3.4 Standard configuration
E.3.4.1 General
This exchange takes place when the manager returns the Associatekesult code of accepted because the manager had presiously been programmed with the documented standard configuration corresponding to the dcv -contig-id sent by the agent. In this case, there is no exchange of configuration information, and the manager and agent base moved into the operating state.